Materials:
Stainless steel material used for manufacturing are in conforming
ASTM F899-2020, EN ISO 7153-1:2016 & EN ISO 21850-1:2020, which is recommended for non-active surgical instruments.
Warning
These products may contain nickel or chromium and should not be used for individuals with a known
sensitivity or allergy to these metals.
Area of Use
Sar sons Surgical Instruments products are designed to perform a specific function such as cutting
grassing clamping dissecting probing retracting aspirating suturing or ligating. They are for use by, a
dental or surgical professional. Products should be used only for the purpose for which they are
designed. The proper dental or surgical technique for the use of any of our products is the responsibility
of the professional. Only trained professionals may use disinfect, clean and sterilize the product.
Inspection
The product must be checked to make sure they work properly before every use.
Inspect all moving parts tips box locks retches and cutting edged. Special attention should be paid to
screws that may have come loose in the ultrasonic cleaner. Damage to the surface, such as scratches
cracks nicks dent etc. as well as bent are indications that should not be used. The products are then to
be repaired or be disposed according to hospital or office procedure. Do not used damaged products.
All products are shipped non sterile and must be decontaminate and sterilize prior to use.
Handling
The products may not be overstressed by twisting or levering, since this can lead to parts becoming
damaged or broken.
Combination with other products and instruments
The products from Sarsons Surgical Instruments may not be combined with products, components and
instruments from other manufacturers under any circumstances. Combinations with products from
other manufacturers may negatively affect the usage of the products and are not allowed since
components used may not be compatible with one another.
Disposal
If the product can no longer be repaired they are to be disposed of according to hospital or office
procedure.
Blow are the recommended sterilization parameters.
STERILIZER |
EXPOSURE TEMPRATURE |
EXPOSURE TIME |
MINIMUM DRY TIME |
Pre-Vacuum (wrapped) |
121o C ( 250o F ) 132o C ( 270o F) 134o C ( 273o F) |
20 minute 4 minutes 3 minutes |
20 minutes 20 minutes 15 minutes |
Pre vacuum (un-wrapped) |
132oC (270oF) |
4 minutes |
|
Gravity Stream (Wrapped) |
132oC (270oF) |
18 minutes |
|