Diethrich Suture 160mm

  • SKU : 07-214-160
  • Availability: IN STOCK
Variations:
Size SKU Description
160mm 07-214-160 Diethrich Suture TC
180mm 07-214-180 Diethrich Suture TC
200mm 07-214-200 Diethrich Suture TC
230mm 07-214-230 Diethrich Suture TC
250mm 07-214-250 Diethrich Suture TC
Product Detail

Materials:

Stainless steel material used for manufacturing are in conforming
ASTM F899-2020, EN ISO 7153-1:2016 & EN ISO 21850-1:2020, which is recommended for non-active surgical instruments.

Warning

These products may contain nickel or chromium and should not be used for individuals with a known

sensitivity or allergy to these metals.

Area of Use

Sar sons Surgical Instruments products are designed to perform a specific function such as cutting

grassing clamping dissecting probing retracting aspirating suturing or ligating. They are for use by, a

dental or surgical professional. Products should be used only for the purpose for which they are

designed. The proper dental or surgical technique for the use of any of our products is the responsibility

of the professional. Only trained professionals may use disinfect, clean and sterilize the product.

Inspection

The product must be checked to make sure they work properly before every use.

Inspect all moving parts tips box locks retches and cutting edged. Special attention should be paid to

screws that may have come loose in the ultrasonic cleaner. Damage to the surface, such as scratches

cracks nicks dent etc. as well as bent are indications that should not be used. The products are then to

be repaired or be disposed according to hospital or office procedure. Do not used damaged products.

All products are shipped non sterile and must be decontaminate and sterilize prior to use.

Handling

The products may not be overstressed by twisting or levering, since this can lead to parts becoming

damaged or broken.

Combination with other products and instruments

The products from Sarsons Surgical Instruments may not be combined with products, components and

instruments from other manufacturers under any circumstances. Combinations with products from

other manufacturers may negatively affect the usage of the products and are not allowed since

components used may not be compatible with one another.

Disposal

If the product can no longer be repaired they are to be disposed of according to hospital or office

procedure.

Additional information

Blow are the recommended sterilization parameters.

 

              

STERILIZER

EXPOSURE TEMPRATURE

EXPOSURE TIME

MINIMUM

DRY TIME

Pre-Vacuum (wrapped)

121C  ( 250o F )

132o C   ( 270F)

134C   ( 273o  F)

20 minute

4 minutes

3 minutes

20 minutes

20 minutes

15 minutes

Pre vacuum (un-wrapped)

132oC   (270oF)

4 minutes

 

Gravity Stream (Wrapped)

132oC   (270oF)

18 minutes